+
button in the window's lower right corner and select the New Environment
menu item. Then specify the Environment's name and select the Pillars
type. It is recommended to use the Antilatency
preset to load the branded reference bar settings.markers per pillar
);pillar length
);Pillar types count
) and proportions for each type (Proportion #
).Pillars vertical offset
);Pillars default FoV
) – applies to all reference bars, but only when creating an Environment. After - only separately for each reference bar;Strict match
). Strict match mode should be enabled when the tracker “sees” strictly the current tracking area (typical use cases). Strict match mode should be disabled if extra reference bars can be visible to the tracker, like in these cases:+ / Add Pillars
button in the window's left side. When you click on the button, it will be underlined with a red line, and the system will offer possible bar types (which were defined by the user during the initial setup).Save
button located in the window's upper right corner. The coordinates and type of each reference bar will be displayed in the Environment Data
table below.Ctrl
button, adjust the angle with the mouse wheel
. By default, reference bars have a 360° angle.Ctrl
button, left-click
on the selected reference bar, and drag in the desired direction.Environment Data
window by entering the required values.Column number | Value |
1 | X axis coordinate |
2 | Y axis coordinate (height above floor level) |
3 | Z axis coordinate |
4 | Bar type (0 — the first type, 1 — the second type, 2 — the third type) |
Apply Data
button.Edit scheme
in the editor and set the value in meters in the Pillars vertical offset
parameter. Now, when you change the additional vertical offset of an individual reference bar, you will need to count from the floor to the base of the reference bar.(constraints)
.Add Constraint
menu button (or Ctrl + Q
).Optimize
button to automatically correct the position.Save
button in the top right corner to save the Environment.Device Network
tab in AntilatencyService.Environments
tab, select your Environment, click ⋮
, then Edit
.Calibrate
button.Begin Collecting
button.Frames count
field).Stop Collecting
button.selected sample
field to select positions that look like tracking errors (incorrect peak points in the tracker path) and delete them (Remove selected
button).Mean sqr error
field. Press the Calibrate
button and then the value of Mean sqr error
should decrease at least by an order of magnitude, and preferably to 0
.Mean sqr error
field is still much greater than 0
, do either:Clear Dataset
and repeat the calibration algorithm.selected sample
line. If no beam hits markers on a reference bar, then the distances between this and other reference bars should be measured again.back
arrow, and the Apply
button in the pop-up window. Then save the Environment changes (Save
button).